GE Precision Healthcare LLC has one OSHA inspection on record in California.

OSHA opened it on Dec 18, 2019 after an accident, in Duarte.

From that inspection OSHA issued 1 citation item: 1 other-than-serious.

OSHA proposed $935 in penalties. After settlements and reviews the total is $750, down 20%.

For context, across all manufacturing employers the median penalty per cited inspection since 2015 is $3,600. GE Precision Healthcare LLC averages $750 over 1 cited inspection in that time.

GE Precision Healthcare LLC: OSHA citation items per year, 2019 to 2019
Citation items per year, by inspection date.

California · latest 90 days of OSHA records

1,133employers cited

Get notified the morning OSHA posts each new citation in California: the employer, the site address, the rules broken and the penalty.

See who gets cited next

Latest cited in California

Checking a subcontractor, a client or your own company? Watch GE Precision Healthcare LLC and get notified the morning OSHA posts a new citation for it.

Watch this company

Inspection history

Inspection 344528518 OPENED DEC 18, 2019Closed

Accident · Partial · Safety · 1 employee at the site · Closed Nov 8, 2022

1500 Duarte Rd., Duarte, CA 91010 · Irradiation Apparatus Manufacturing · OSHA's record · Phone and website on Google Maps

Incident on Nov 26, 2019: Employee amputates finger when pulled into drive chain

At 2:30 p.m. on November 26, 2019, an employee who worked as an industrial engineer for a medical radiological device manufacturer was troubleshooting a Revolution Model XKD X-ray Table. A malfunction was occurring in the vertical function of the table. The issue appeared to be a motor that was preventing the table from going up and down. The employee removed the cover to the motor compartment and attempted to engage the drive chain. The employee was operating the foot pedal and reached into the compartment with his gloved right hand to touch the chain to see if there was tension. As he reached inside the compartment, the motor engaged causing the chain to move. The chain caught his pinky finger at the base of the finger. The employee sustained an open fracture and amputation to his finger, requiring hospitalization.

Citation and ruleTypeProposedCurrentIssuedStatus
Citation 1, item 1
29 CFR 3314(D)
Other-than-serious$935$750May 20, 2020Final

The rules behind these citations

Current text from the Code of Federal Regulations. Some citations were issued under older wording.

Other irradiation apparatus manufacturing employers in California

Alert subscribers get notified of each newly cited employer in California the morning it is posted. See the alert plans.

Source: U.S. Department of Labor, OSHA enforcement data. A citation is OSHA's allegation; employers can contest it, and the status column shows where each item stands. Something wrong on this page? Tell us.