OSHA SearchOSHA records by employer

SDS Dental Inc.: OSHA inspections and violations

Pompano Beach, Florida · Medical, Dental, and Hospital Equipment and Supplies Merchant Wholesalers · NAICS 423450

1inspections
5citation items
$17,076current penalties
Nov 15, 2017last inspection

SDS Dental Inc. has one OSHA inspection on record in Florida.

OSHA opened it on Nov 15, 2017 on a referral, in Pompano Beach.

From that inspection OSHA issued 5 citation items: 4 serious and 1 other-than-serious.

OSHA proposed $24,392 in penalties. After settlements and reviews the total is $17,076, down 30%.

Most often, OSHA cited 1910.335, safeguards for personnel protection (2 items).

For context, across all wholesale employers the median penalty per cited inspection since 2015 is $3,425. SDS Dental Inc. averages $17,076 over 1 cited inspection in that time.

SDS Dental Inc. also files yearly injury logs with OSHA, summarized below.

SDS Dental Inc.: OSHA citation items per year, 2017 to 2017
Citation items per year, by inspection date.

Florida · latest 90 days of OSHA records

320employers cited

Get notified the morning OSHA posts each new citation in Florida: the employer, the site address, the rules broken and the penalty.

See who gets cited next

Latest cited in Florida

Checking a subcontractor, a client or your own company? Watch SDS Dental Inc. and get notified the morning OSHA posts a new citation for it.

Watch this company

Inspection history

Inspection 342775111 OPENED NOV 15, 2017Closed

Referral · Partial · Safety · 18 employees at the site · Closed Apr 13, 2023

1280 Sw 27th Avenue, Pompano Beach, FL 33069 · Medical, Dental, and Hospital Equipment and Supplies Merchant Wholesalers · OSHA's record · Phone and website on Google Maps

Incident on Nov 14, 2017: Employee drills into electrical panel and is burned by an ar

On November 14, 2017, an employee was drilling a hole with a hand-held battery operated portable electrical drill into an energized electrical panel that contained energized components. The employee was burned on his head, chest, and hands from an arc flash.

Citation and ruleTypeProposedCurrentIssuedStatus
Citation 1, item 1
29 CFR 1910.333(a)(1) · Working near live electrical parts
Serious$5,174$3,622Apr 13, 2018Final
Citation 1, item 2
29 CFR 1910.333(c)(2) · Working near live electrical parts
Serious$5,174$3,622Apr 13, 2018Final
Citation 1, item 3
29 CFR 1910.335(a)(1)(i) · Safeguards for personnel protection
Serious$5,174$3,622Apr 13, 2018Final
Citation 1, item 4
29 CFR 1910.335(a)(2)(i) · Safeguards for personnel protection
Serious$5,174$3,622Apr 13, 2018Final
Citation 2, item 1
29 CFR 1904.39(a)(2) · Reporting deaths and severe injuries
Other-than-serious$3,696$2,588Apr 13, 2018Final

Injury and illness logs

From the Form 300A summaries employers file with OSHA each year. Rate = recordable cases per 100 full-time workers.

In 2017, the Pompano Beach site logged 1 recordable injury or illness among about 18 workers: 11.6 per 100 full-time workers, against a median of 1.1 for its industry. Those cases cost 12 days away from work.

Site and yearWorkersCasesDays-away casesRateIndustry median
Summit Dental Systems
Pompano Beach · 2017
18 1111.61.1

The rules behind these citations

Current text from the Code of Federal Regulations. Some citations were issued under older wording.

29 CFR 1904.39 · Reporting deaths and severe injuries

29 CFR 1904.39(a)(2)

(2) Within twenty-four (24) hours after the in-patient hospitalization of one or more employees or an employee's amputation or an employee's loss of an eye, as a result of a work-related incident, you must report the in-patient hospitalization, amputation, or loss of an eye to OSHA.

29 CFR 1910.333 · Working near live electrical parts

29 CFR 1910.333(a)(1)

(1) Deenergized parts. Live parts to which an employee may be exposed shall be deenergized before the employee works on or near them, unless the employer can demonstrate that deenergizing introduces additional or increased hazards or is infeasible due to equipment design or operational limitations. Live parts that operate at less than 50 volts to ground need not be deenergized if there will be no increased exposure to electrical burns or to explosion due to electric arcs.

29 CFR 1910.333(c)(2)

(2) Work on energized equipment. Only qualified persons may work on electric circuit parts or equipment that have not been deenergized under the procedures of paragraph (b) of this section. Such persons shall be capable of working safely on energized circuits and shall be familiar with the proper use of special precautionary techniques, personal protective equipment, insulating and shielding materials, and insulated tools.

29 CFR 1910.335 · Safeguards for personnel protection

Other medical, dental, and hospital equipment and supplies merchant wholesalers employers in Florida

Alert subscribers get notified of each newly cited employer in Florida the morning it is posted. See the alert plans.

Source: U.S. Department of Labor, OSHA enforcement data. A citation is OSHA's allegation; employers can contest it, and the status column shows where each item stands. Something wrong on this page? Tell us.